Conductivity to TDS Calculator

Every conductivity meter can be told to display TDS, and every one of them needs a conversion factor to do it. Get that factor wrong and the number on the display is wrong — silently, because the meter has no way to check it. This calculator shows the arithmetic and, more usefully, which factor belongs to which application.

TDS from conductivity

Why the same water reads two different TDS values

Conductivity measures how well water carries an electric current, which depends on dissolved ions. TDS estimates total dissolved mass from that conductivity using a multiplier. The multiplier is not a physical constant — different salts carry current differently, so the same water can justify 0.5 or 0.7 depending on what is dissolved in it.

The practical consequence: a conductivity meter set to TDS is a calibrated guess. It is fine for trend watching and screening, and wrong for compliance reporting. Where a permit limit exists, report conductivity in µS/cm and let the limit be written in conductivity terms.

Conversion factors by application

ApplicationTypical factorBasisCaution
Potable / drinking water0.5 – 0.7Mostly Ca₂/Mg₂ saltsWide range; use the local authority value if one exists
Seawater / brine0.5NaCl dominatedAbove 20 mS/cm consider salinity instead of TDS
Hydroponics / fertigation0.7Mixed fertiliser salts0.7 (“442”) is the horticultural convention
Boiler feedwater—Very low conductivityTDS is not meaningful below ~10 µS/cm; report conductivity
Industrial rinse / reclaimsite-specificProcess dependentMeasure gravimetrically, then set your own factor
Laboratory / QC1.0 or gravimetricKnown standardsDo not derive TDS when you already have the mass

Reading TDS on your own instrument

Most panel meters expose a factor setting, usually labelled K or TDS factor, with selectable values like 0.5, 0.6 and 0.7. The instrument multiplies the conductivity reading by that number and displays the product as mg/L. The default is often 0.5 because NaCl is the calibration standard, not because 0.5 is right for your water.

If your process is mostly a single salt — brine, a rinse stream, a specific fertiliser mix — you can determine your own factor in one afternoon: measure conductivity, take a gravimetric TDS on the same sample, and divide. That site-specific factor beats any default.

Temperature and compensation

Conductivity changes about 2% per °C in most aqueous solutions. An uncompensated meter reading 1000µS/cm at 15°C will read roughly 1200µS/cm at 25°C. Automatic temperature compensation normalises to a reference temperature, usually 25°C, so measurements stay comparable across seasons. For TDS derived from conductivity, this matters just as much as the reading itself.

Instruments for this measurement

3-in-1 pen tester: conductivity, TDS, salinity

Pocket size · field inspection

Handheld conductivity / TDS / salinity tester

Wide spectrum to 400 mS/cm · IP65

Benchtop conductivity / TDS / salinity meter

0.001 µS/cm resolution · 256 data storage

98×98 mm panel conductivity transmitter

Compact cut-out · conductivity, TDS, temperature

144×144 mm panel conductivity monitor

Cell constants K=0.01–10 · 4–20 mS/cm

Inductive non-contact conductivity transmitter

K=2.7 · to 2000 mS/cm · no electrodes

Models shown are representative families, not a complete catalogue. Send us your application, water analysis and installation constraints and we will match instruments to it.

Frequently asked

What is the conversion factor from conductivity to TDS?
Between 0.5 and 0.7 for most applications. 0.5 corresponds to the NaCl standard and suits brine and drinking water; 0.7 (the “442” factor) is the horticultural convention. Anything outside that band should be measured gravimetrically and set per site.
Is TDS the same as conductivity?
No. Conductivity is a measured electrical property in µS/cm. TDS is a calculated mass estimate in mg/L derived from it. The relationship depends on what salts are present, which is why a single factor cannot be universally correct.
Why does my TDS change with temperature if I compensate conductivity?
Compensation normalises the conductivity reading to a reference temperature. Because TDS is a fixed multiple of that reading, the compensation carries through — but the underlying ion mobility also shifts with temperature, so the factor itself is mildly temperature-dependent. For most process work, 25°C compensation is sufficient.
Should I use TDS or conductivity for a discharge permit?
Use whichever the permit writes in. If the limit is in mg/L and you report conductivity, the conversion error becomes your problem. Ask the authority to state the limit in the unit you can measure most accurately.
Can I report TDS for compliance?
Only if the permit explicitly accepts a calculated value. Gravimetric TDS (EPA 160.2 / SM 2540 D) is the reference method. Calculated TDS is a screening estimate, and the method note should say so.

This page provides calculation guidance and a screening cross-check. It is not a substitute for the measurement method named in your permit, and values shown are typical ranges rather than guarantees for a particular instrument.

Cite this tool / 引用此工具

If you use this result in a report, specification or client document, please cite it as below. It takes one line and lets engineers find us. 如果您把计算结果用于报告、规格书或客户文件,请按下列方式引用。

Jiuyuhub. "Conductivity to TDS Calculator". Jiuyuhub, https://jiuyuhub.com/conductivity-tds-converter.html