Conductivity to TDS Calculator
Every conductivity meter can be told to display TDS, and every one of them needs a conversion factor to do it. Get that factor wrong and the number on the display is wrong — silently, because the meter has no way to check it. This calculator shows the arithmetic and, more usefully, which factor belongs to which application.
TDS from conductivity
Result
Why the same water reads two different TDS values
Conductivity measures how well water carries an electric current, which depends on dissolved ions. TDS estimates total dissolved mass from that conductivity using a multiplier. The multiplier is not a physical constant — different salts carry current differently, so the same water can justify 0.5 or 0.7 depending on what is dissolved in it.
The practical consequence: a conductivity meter set to TDS is a calibrated guess. It is fine for trend watching and screening, and wrong for compliance reporting. Where a permit limit exists, report conductivity in µS/cm and let the limit be written in conductivity terms.
Conversion factors by application
| Application | Typical factor | Basis | Caution |
|---|---|---|---|
| Potable / drinking water | 0.5 – 0.7 | Mostly Ca₂/Mg₂ salts | Wide range; use the local authority value if one exists |
| Seawater / brine | 0.5 | NaCl dominated | Above 20 mS/cm consider salinity instead of TDS |
| Hydroponics / fertigation | 0.7 | Mixed fertiliser salts | 0.7 (“442”) is the horticultural convention |
| Boiler feedwater | — | Very low conductivity | TDS is not meaningful below ~10 µS/cm; report conductivity |
| Industrial rinse / reclaim | site-specific | Process dependent | Measure gravimetrically, then set your own factor |
| Laboratory / QC | 1.0 or gravimetric | Known standards | Do not derive TDS when you already have the mass |
Reading TDS on your own instrument
Most panel meters expose a factor setting, usually labelled K or TDS factor, with selectable values like 0.5, 0.6 and 0.7. The instrument multiplies the conductivity reading by that number and displays the product as mg/L. The default is often 0.5 because NaCl is the calibration standard, not because 0.5 is right for your water.
If your process is mostly a single salt — brine, a rinse stream, a specific fertiliser mix — you can determine your own factor in one afternoon: measure conductivity, take a gravimetric TDS on the same sample, and divide. That site-specific factor beats any default.
Temperature and compensation
Conductivity changes about 2% per °C in most aqueous solutions. An uncompensated meter reading 1000µS/cm at 15°C will read roughly 1200µS/cm at 25°C. Automatic temperature compensation normalises to a reference temperature, usually 25°C, so measurements stay comparable across seasons. For TDS derived from conductivity, this matters just as much as the reading itself.
Instruments for this measurement
Models shown are representative families, not a complete catalogue. Send us your application, water analysis and installation constraints and we will match instruments to it.
Frequently asked
What is the conversion factor from conductivity to TDS?
Is TDS the same as conductivity?
Why does my TDS change with temperature if I compensate conductivity?
Should I use TDS or conductivity for a discharge permit?
Can I report TDS for compliance?
This page provides calculation guidance and a screening cross-check. It is not a substitute for the measurement method named in your permit, and values shown are typical ranges rather than guarantees for a particular instrument.
Cite this tool / 引用此工具
If you use this result in a report, specification or client document, please cite it as below. It takes one line and lets engineers find us. 如果您把计算结果用于报告、规格书或客户文件,请按下列方式引用。
Jiuyuhub. "Conductivity to TDS Calculator". Jiuyuhub, https://jiuyuhub.com/conductivity-tds-converter.html
Licensed for internal and commercial technical use (CC BY 4.0). 内容可用于内部及商业技术用途。 Questions? Ask our engineers