Turbidity to TSS Converter

Turbidity meters read NTU. Discharge permits and most laboratory methods are written in mg/L. There is no universal number that converts one to the other, because the correlation depends on particle size, shape, colour and refractive index of the solids in your water. What this calculator does is make the factor explicit — and show you how to determine your own.

TSS from turbidity

Why there is no universal NTU to mg/L factor

A nephelometer measures light scattered at 90 degrees. That response depends on how large, how shaped and how refractive the particles are. A gram of fine clay scatters very differently from a gram of sand, at the same mass. This is why regulatory methods keep turbidity and suspended solids as separate parameters rather than treating one as a proxy for the other.

Published benchmark correlations exist — roughly 1.0 to 2.0 mg/L per NTU depending on water type — and they are useful for estimating, not for reporting.

Benchmark correlations by water type

Water typeBenchmark factorTypical NTUTurbidity meters (NTU)
Treated drinking water1.0 – 1.50.3 – 20.01 – 10 NTU, low range
Raw surface water1.5 – 2.05 – 1000.01–100 NTU with switch
Post-sedimentation basin1.0 – 1.51 – 100.01–10 NTU, low range
Final effluent (reclaimed)1.0 – 2.01 – 200.01–100 NTU
Stormwater / runoffsite-specific20 – 5000–1000 NTU dual range
Industrial high solidssite-specific500 – 5000Switched range + dilution

Determining your site-specific factor

The defensible method takes one afternoon. Collect a representative sample from the exact discharge point. Measure NTU in the field with a calibrated meter. Send a second portion to a laboratory for certified gravimetric TSS by EPA 160.2 or Standard Method 2540 D. Divide the laboratory result by your field reading, and you have a factor that is valid for that water.

Do this across a few conditions — wet weather, dry weather, first flush — and if the factor varies, you have learned something important: a single NTU reading is not a reliable TSS estimate for your discharge.

What turbidity range to specify

Range selection is driven by the concentration you need to resolve, not by the maximum possible. Potable water targets below 1 NTU need a low-range sensor with 0.01 NTU resolution. Wastewater clarification typically runs 5–50 NTU, so a 0–100 NTU range is appropriate. High-solid industrial streams can reach thousands of NTU and need a switched or dual-range instrument with automatic dilution.

Instruments for this measurement

Panel turbidity monitor, switchable window

0–20 / 0–200 / 0–2000 / 0–4000 NTU

Panel turbidity monitor, low range

0.01–10 NTU · for filtered water

Models shown are representative families, not a complete catalogue. Send us your application, water analysis and installation constraints and we will match instruments to it.

Frequently asked

How do you convert NTU to mg/L of TSS?
Multiply the NTU reading by a correlation factor. For treated drinking water 1.0–1.5 is typical; for raw surface water 1.5–2.0. For compliance, derive the factor from a paired field and laboratory sample rather than using a published average.
Is 1 NTU equal to 1 mg/L?
No. A common rule of thumb used in some older references is 1 mg/L ≈ 3 NTU, but that only holds for one particle type under one set of conditions. It is not transferable between water bodies or between treatment stages.
Can a turbidity meter replace a TSS measurement for permit reporting?
No. Turbidity is a different physical measurement, and regulatory methods specify gravimetric or filtration-based TSS. A turbidity meter can verify that a clarifier is working and can trigger an alarm, but the reported TSS has to come from a method the authority accepts.
Why does my turbidity reading fall after filtration but the laboratory TSS barely changes?
Likely particle size distribution. A filter can remove large, highly scattering particles while leaving fine colloids that contribute little to 90-degree scatter but carry real mass. That divergence is a signal that the correlation factor no longer applies to the filtered water.
Does NTU depend on the instrument light source?
Yes, and that is why two standards exist. NTU follows US EPA Method 180.1 (white light); FNU follows ISO 7027 (infrared). They agree on the formazin calibration standard but can differ on real samples. Mixing NTU and FNU readings in one dataset will produce drift that is not the water’s fault.

This page provides calculation guidance and a screening cross-check. It is not a substitute for the measurement method named in your permit, and values shown are typical ranges rather than guarantees for a particular instrument.

Cite this tool / 引用此工具

If you use this result in a report, specification or client document, please cite it as below. It takes one line and lets engineers find us. 如果您把计算结果用于报告、规格书或客户文件,请按下列方式引用。

Jiuyuhub. "Turbidity to TSS Converter". Jiuyuhub, https://jiuyuhub.com/turbidity-tss-ntu-converter.html

Licensed for internal and commercial technical use (CC BY 4.0). 内容可用于内部及商业技术用途。 Questions? Ask our engineers