CON3000C 98×98 compact conductivity monitor for TDS/temperature. For power plant cooling water

98×98 compact conductivity monitor for TDS/temperature. For power plant cooling water, make-up water, saturated water, and seawater distillation.
✓ Transparent sourcing — factory-direct price, no markup
Request a Quote WhatsApp Us

Product Specifications

Measurement Range (set by electrode constant)Conductivity: K=0.01: 0~20 uS·cm-1; K=0.1: 0~200 uS·cm-1; K=1.0: 0~4000 uS·cm-1; K=10.0: 0~40.0 mS·cm-1; K=30.0: 0~100.0 mS·cm-1
TDSK=0.01: 0~10 mg/L; K=0.1: 0~100 mg/L; K=1.0: 0~2000 mg/L; K=10.0: 0~20000 mg/L; K=30.0: 0~50000 mg/L
Temperature0~150℃ (depending on electrode configuration)
ResolutionConductivity: 0.01 uS·cm-1; 0.01 mS·cm-1; TDS: 0.01 mg/L; 0.01 g/L; Temperature: 0.1℃
Basic ErrorConductivity/TDS: ±1.0% F.S. ±1 digit; Temperature: ±0.5℃
Auto/Manual Temp Compensation Range0~150℃ (reference 25℃)
Auto Temp Compensation Error±0.3% F.S.
Electronic Unit Stability±0.1% F.S. ±1 digit/24h
Electronic Unit Repeatability Error≤0.1% F.S. ±1 digit
Electronic Unit Alarm Error±1% F.S.
Electronic Unit Output Current Error±1% F.S.
Signal Output0~10 mA (load resistance <1.5 KΩ); 4~20 mA (load resistance <750 Ω); 20~4 mA (load resistance <750 Ω)
2 Relay Control Contacts3A 240 VAC, 3A 28 VDC or 120 VAC
Power Supply (optional)85–265 VAC±10%, 50±1 Hz, power ≤3W; 9–36 VDC, power ≤3W
External Dimensions98×98×130mm
Installation MethodPanel mount (embedded)
Panel Cutout93×93mm
Instrument Weight0.6kg
Working Environment-10~60℃; relative humidity ≤90%; no strong magnetic field interference other than the earth's magnetic field
ApplicationsWater treatment, environmental monitoring, laboratories, food & beverage, aquaculture
CertificationsManufactured under ISO 9001 QMS · CE/RoHS/FDA compliance varies by model (certificates available upon request)

概述

CON3000C 是一款 98 × 98 mm 紧凑型面板电导率监控仪,通过一支电极同时输出电导率、溶解性总固体(TDS)和温度三个数值。外形尺寸 98 × 98 × 130 mm、开孔 93 × 93 mm、整机重 0.6 kg,适合装在标准 144 × 144 mm 仪表放不进去的控制柜与成套撬装设备中。

量程由电极常数决定,而不是由变送器决定。本机支持 K=0.01(0–20 uS/cm)、K=0.1、K=1.0、K=10.0 到 K=30.0(0–100.0 mS/cm)共五档常数,TDS 随常数对应从 0–10 mg/L 到 50 000 mg/L。基本误差:电导率与 TDS 为满量程的 ±1.0% ±1 个字,温度为 ±0.5 °C。

应用场景

选型建议

  1. 订货前先定电极常数。变送器只能读出电极所允许的量程:超纯水与锅炉给水选 K=0.01,一般工艺水选 K=1.0,浓盐水与海水蒸馏选 K=10.0 或 K=30.0。
  2. 温度补偿范围 0–150 °C,以 25 °C 为参比,自动补偿精度为满量程的 ±0.3%。只有当实验室方法明确要求时,才切换到手动补偿。
  3. 注意继电器数量。本机提供两组控制触点,容量 3 A/240 VAC 或 3 A/28 VDC —— 够用一路报警加一路阀门,但不足以做三段式排污程序。
  4. 核对回路阻抗。0–10 mA 输出可驱动最大 1.5 kΩ 负载,而 4–20 mA 与 20–4 mA 输出要求负载低于 750 Ω。

常见问题

CON3000C 与 144 × 144 mm 电导率仪有什么区别?

主要是面板空间和继电器数量。CON3000C 开孔 93 × 93 mm、提供两组继电器触点;大表壳提供三组且显示面积更大。测量性能基本相当。

TDS 是直接测出来的吗?

不是。TDS 由电导率读数按换算系数计算得出,因此其准确度取决于该系数是否贴合您水样中实际溶解固体的成分。

电极能耐多高温度?

最高 150 °C,具体取决于所选电极配置。订货时务必确认电极本体的耐温等级,因为变送器与电极的限值并不相同。

需要什么电源?

85–265 VAC(50 Hz)或 9–36 VDC 均可,两种供电功耗都不超过 3 W。直流方案适合太阳能供电和移动式装置。

两次校准之间读数稳定吗?

电子单元 24 小时稳定性为满量程的 ±0.1% ±1 个字,重复性在 0.1% 以内。实际使用中的漂移多来自电极结垢,因此清洗周期比校准频率更关键。