| Measurement Range | DO: 0~25.00 mg/L; Temperature: 0~60℃ |
|---|---|
| Resolution | 0.01 mg/L |
| Measurement Error | DO: ±1% F.S.; Temperature: ±0.5℃ |
| Automatic Temperature Compensation | 0~60℃ |
| Electrode Residual Signal | < 1‰ |
| Response Time (90% final value) | < 60s at 25°C; < 30s at 35°C |
| Stability | Drift < 2% F.S. per week at constant pressure and temperature |
| Signal Output | 0~10 mA (load resistance < 1.5K Ω); 4~20 mA (load resistance < 750 Ω); 20~4 mA (load resistance < 750 Ω) |
| 2 Relay Control Contacts | 3A 240 VAC, 3A 28 VDC or 120 VAC |
| Power Supply | 85~265 VAC±10%, 50±1 Hz, power ≤3W; 9~36 VDC, power ≤3W (optional) |
| External Dimensions | 98×98×130mm |
| Installation Method | Panel mount (embedded) |
| Panel Cutout | 93×93mm |
| Instrument Weight | 0.6kg |
| Working Environment | Ambient temp -10~60°C; relative humidity ≤90%; no strong magnetic field interference (except earth magnetic field) |
| Applications | Water treatment, environmental monitoring, laboratories, food & beverage, aquaculture |
| Certifications | Manufactured under ISO 9001 QMS · CE/RoHS/FDA compliance varies by model (certificates available upon request) |
Ang DO3000M ay ang compact na 98 × 98 mm na miyembro ng dissolved oxygen monitor family, na binuo sa paligid ng parehong membrane-covered amperometric electrode technology gaya ng mas malalaking panel model. Sumusukat ito ng dissolved oxygen mula 0 hanggang 25.00 mg/L na may resolution na 0.01 mg/L at basic error na ±1% ng full scale, kasama ang temperatura mula 0 hanggang 60 °C na may automatic compensation sa parehong range. Sa 98 × 98 × 130 mm at 0.6 kg na may 93 × 93 mm panel cut-out, kasya ito sa mga cabinet na hindi kayang tanggapin ang 144 × 144 mm na instrumento.
Ang pagtugon at katatagan ay angkop sa aeration control: naabot ng electrode ang 90% ng final value sa loob ng wala pang 60 segundo sa 25 °C at wala pang 30 segundo sa 35 °C, at ang drift ay nananatiling mas mababa sa 2% ng full scale bawat linggo sa pare-parehong pressure at temperatura. Ang signal output ay maaaring itakda sa 0–10 mA (load hanggang 1.5 kΩ) o sa 4–20 at 20–4 mA (load na mas mababa sa 750 Ω). Dalawang relay na may rating na 3 A sa 240 VAC (o 3 A sa 28 VDC) ang nagmamaneho ng mga alarm at isang blower stage, at ang metro ay tumatakbo sa 85–265 VAC o, bilang option, 9–36 VDC na hindi lalampas sa 3 W.
Paano naiiba ang DO3000M sa 144 × 144 mm na DO6000M?
Pareho silang gumagamit ng membrane electrode at sumusukat ng 0–25.00 mg/L sa ±1% ng full scale. Ang DO3000M ay ang compact 98 × 98 mm na bersyon na may 93 × 93 mm cut-out at dalawang relay; ang DO6000M ay nagdaragdag ng ikatlong relay, wall mounting, at mas mahigpit na temperature accuracy.
Ano ang range at accuracy na sinusukat nito?
Dissolved oxygen mula 0 hanggang 25.00 mg/L na may basic error na ±1% ng full scale at 0.01 mg/L resolution, kasama ang temperatura mula 0 hanggang 60 °C na may automatic compensation.
Gaano ito kabilis tumugon?
Naaabot ng electrode ang 90% ng final value sa loob ng wala pang 60 segundo sa 25 °C at wala pang 30 segundo sa 35 °C, sapat na mabilis para sa blower control na sumusubaybay sa nagbabagong oxygen demand.
Paano ito pinapagana?
Sa pamamagitan ng 85–265 VAC sa 50 Hz, o bilang option na 9–36 VDC, na kumukonsumo ng hindi hihigit sa 3 W sa alinmang supply - angkop ang DC option para sa solar o battery-backed remote panel.
Anong maintenance ang kailangan ng electrode?
Ang membrane at electrolyte ay consumable na may limitadong working life. Sundin ang electrode manual para sa filling, storage, at cleaning, palitan ang membrane kapag bumagal ang pagtugon, at i-calibrate laban sa reference tulad ng water-saturated air sa nakapirming schedule.
This product is sourced factory-direct from a certified Chinese manufacturer. You pay the factory price — no markup. Our income comes from the export tax rebate, not from your bill.