| Measurement Range | 0.0 uS/cm~20.00 mS/cm |
|---|---|
| Resolution | 0.1 uS/cm~0.01 mS/cm |
| Accuracy | ±1%F.S. |
| Calibration | 1 point (any point within full range) |
| Electrode | Replaceable K=1 graphite electrode head |
| Temperature Measuring Range | 0~100.0℃/ 32~212℉ |
| Operating Temperature Range | 0~60.0℃/ 32~140℉ |
| TDS Measuring Range | 0.0mg/L~20.0g/L |
| TDS Coefficient | 0.40~1.00 (factory default 0.50) |
| Salinity Measuring Range | 0.0mg/L~13.0g/L |
| Salinity Coefficient | 0.65 |
| Display | 20×30mm multi-line LCD display |
| Temperature Compensation | ATC automatic / MTC manual |
| Protection Rating | IP67 |
| Power Supply | 1 × 1.5V AAA battery |
| Auto Backlight | 1 minute |
| Auto Power Off | 10 minutes |
| Dimensions | 180×40×45mm |
| Weight | 90g |
| Applications | Water treatment, environmental monitoring, laboratories, food & beverage, aquaculture |
| Certifications | Manufactured under ISO 9001 QMS · CE/RoHS/FDA compliance varies by model (certificates available upon request) |
CON100 是一款笔式便携检测仪,用同一支电极、对同一样品读取三个数值:电导率、溶解性总固体(TDS)和盐度。电导率为实测值;TDS 与盐度则由该读数经换算系数计算得出——因此仪表允许你在 0.40 至 1.00 之间任意设定 TDS 系数,而不是固定在某一个数值上。
它是一台抽检仪表,不是在线监测仪。电极内置在机身中,所以使用方式是取样、浸入、读数、冲洗。如果你需要把数值连续记录进 PLC 或 SCADA 系统,那应该选在线电导率变送器。
CON100 能测海水吗?
不能。盐度上限为 13.0 g/L,电导率上限为 20.00 mS/cm。天然海水约为 35 g/L,远超该范围,需要专用的盐度计或更大量程的电极。
多久需要校准一次?
校准为一点校准,可在全量程内任意点进行。日常现场使用下,每周用接近你常用读数的标准液核查一次即可;让标准液贴近实际测量值,比总是在 1413 uS/cm 校准更准确。
电极可以更换吗?
可以。K=1 石墨电极头是易耗件,无需返厂即可更换。
该用 ATC 还是 MTC?
现场使用保持 ATC。只有在需要对接以固定 25 °C 报出结果的实验室方法时,才使用 MTC。
电池能用多久?
使用一节 1.5 V 七号电池。背光 1 分钟后自动关闭,仪表在 10 分钟无操作后自动关机;这两个时间都不可调整。